2. About Godox Radio System
●
7.
Press the Measuring Button 6.
The measured value (F-stop) will be displayed. Press the Measuring Button
6
again to re a ash and measure until the desired F-stop is displayed.
The accumulated measured value (F-stop) and the cumulative count will be
displayed.
If you want to adjust the output power of each ash separately, go to Flash Power
Control Screen to make a measurement and adjustment.
In the Flash Power
Control Screen, the Multi (cumulative) ash measurement is not
available. (
P16)
NOTICE
● In case of the following, please follow “Cord Multi (Cumu.) Flash Mode” ( ).
• When ring the ash, if the ash brightness is lower than the ambient light, the meter
may fail to detect the light.
• Rapid start uorescent lamps and special lighting are sometimes mistaken for ash
and accidentally measured.
• Even if the ash does not re, when a sudden light change occurs in the light receptor,
measurement may be made.
• The waveform of a ash bulb has a slight slope and there is a possibility that the light
meter cannot recognize the ash bulb.
● The EV scale cannot be displayed in the Radio Triggering Multi (Cumu.) Flash mode.
NOTE
● For red-eye prevention and auto-ash light adjustment, some devices can pre-ash
before main ash burst. With the normal setting, the light meter will measure the pre-
ash bursts and not the main ash burst. To take a successful reading, activate the pre-
ash feature in the Tool Box. Please refer to the L-858D Operating Manual.
● For radio CH frequencies, refer to "5. Radio Channel Frequencies". (
P48)
[Measuring Screen]
Measured Value
(F-stop)
Cumulative Count