44
2. About Godox Radio System
4.
Enter the "Reference" of 0.1 to 0.9 by touching the numeric value.
The t value can be set in steps of 0.1 at a range of 0.1 to 0.9.
The rst "0." is xed. Enter the rst digit decimal only. (To set "0.1", enter "1".)
5.
Touch [OK] Button.
The setting is entered, and the display returns to the Measuring Screen.
Touch the [Cancel] Button to return to the Measuring Screen without making the modication.
NOTE
Two rules apply to the reference ash duration
time.
t0.5 = Effective ash duration time
t0.1 = Total ash duration time
After ash ring, the time at which the
maximum intensity drops by half is called
"t0.5". The time at which the maximum intensity
drops to 1/10 is called "t0.1".
Generally, "t0.5" is called the ash duration
time.
[Flash Duration Analysis
t Value Screen]
[OK] Button [Cancel] Button
The input
numeric
value is
displayed.
[Measuring Screen]
t
0.1
t
0.2
t
0.3
t
0.4
t
0.5
t
0.6
t
0.7
t
0.8
t
0.9
[Flash Duration Analysis
t Value Screen]