EasyManua.ls Logo

Selecta 4002429 - General Information; Packing List; Technical Specifications

Selecta 4002429
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MANUAL DE INSTRUCCIONES CODIGO 80342 REV O 11/2021 Pag.: 33
J.P. SELECTA s.a.u. Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España Tel 34 937 700 877 Fax 34 937 702 362
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com
1. General Information
• Handle the package with care. Unpack it and check that the content matches what is indicated in the “Pac-
king list” section. If any damaged component or the absence of any is observed, notify the distributor quickly.
• Do not install or use the equipment without rst reading this instruction manual. These instructions are an
inseparable part of the equipment and must be available to all its users.
• Consult any doubt to the technical service of J.P. SELECTA, s.a.u.
• Do not use the equipment in explosive, ammable or corrosive atmospheres.
2. Packing List
Item 150 L 80L 20L
4002423 4002424 4002429
Hoses 3/4’ (Steam / Water / Drain) x3 46143
1x Press lids for jar lids 34370
1x Drill jar lids 12039
3.TechnicalSpecications
4002423 4002424 4002429
Item
150L
80L 20L
Network voltage: III 400V I 230V I 230V
Electric power: (W) 7500 4000 2800
Maximum intensity per phase: (A) 12 18 12
Fuses: (fast, type H, dimensions: 10x38mm) 32A 20A 20A
Network connection cable: R, S, T, Neutral, Earth (section 2.5mm
2
) L/N/PE (section 4mm
2
) L/N/PE (section 1.5mm
2
)
Weight (Net Kg): 250 125 50
Useful dimensions (cm Ø x depth) Ø50x70 Ø40x60 Ø25x40
Ext measurement: (cm Depth x Width x Height) 95x120x80 72x58x117 70x38x38
Nominal Volume: (Liters) 150 80 20
Maximum load: (Kg) (Metal) 50 20 10
Weight per support (Kg / cm2) Empty 2.5 2.5 2.5
Weight per support (Kg / cm2) Load + Water 2.5 2.5 2.5
Total space occupied + doors + (optional crane) Add 1.2 m to height Add 40cm to height *Does not adapt
Minimum external water pressure (relative bar) 2
Volume of water per cycle (Liters) 25 20 15
Water quality * Soft water (see classication table)
Thermal energy transmitted to the environment: (W / h) 525 600 375
External water inlet connection 3/4’’
Acoustic Level: Below 70 dBA
Operating regime: Continuous with intervals of 20 minutes between each cycle.
Environmental conditions: Indoor use. Altitude up to 2000m
Ambient temperature between Ambient temperature 5ºC and 40ºC.
Maximum relative humidity 80% up to 31ºC, linearly up to 50% relative humidity at 40ºC
Overvoltage level: Category II
Degree of contamination: 2
Material in contact with steam: Stainless steel, Copper, Teon, Brass
Chamber material: Stainless Steel
Pipes material: Copper, Brass

Related product manuals