Programmable lethality values (F
0
Custom)
F
0
Custom option allows the creation of a lethality table according to the
reference microorganism that the client wishes, to obtain the total lethality
value of the programmed heating treatment.
Note: When editing the program, you must select “F
0
Custom” type to use.
The partial lethality values are obtained by applying the formula:
L = 10 ^ ((T-Tx) / z)
The user must enter the Z and Tx values (unique to each microorganism)
and the custom lethality table will be created. The denition of these
parameters is:
Z: Temperature to reduce a reference microbial population by a loga-
rithmic power.
Tx: Reference temperature.
Most common reference values:
Download logs
All logs are stored in the autoclave’s memory. The logs can be downloaded to
the USB Pendrive. Logs are removed from the autoclave, when downloaded.
See section 10 to present the data in graphic form.
Information
It allows to install the autoclave software updates.
Check with service@jpselecta.es to nd out about the available software
versions.
Referencebacteria Tx(°C) Z(°C)
Listeria monocytogenes 70
7,5
Staphylococcus aureus 70
Salmonella typhimurium 70
Yersinia enterocolitica 70
E. Coli 70
C jejuni 70
Spore-forming bacillus in acid medium 93,3 8,9
Clostridium Botulinum proteolítico 121,1 10