B. QUICK GUIDE
1. TO CALIBRATE FLAT SHEET PROBE/FORMED CAN PROBE.
A) Attach crocodile clip to test sample but make sure probe is not
in contact.
B) Enter set-up mode by: pressing CALIBRATE and LACQUER
NUMBER together. (PASS should appear in large display, if
not press END and try set-up mode again) then enter
PASSWORD. (9963 if unchanged).
C) Press CALIBRATE. (CP will appear in the small display).
D) Press ENTER. (The large display will show 0.00)
E) Press END twice.
2. TO SET-UP A LACQUER TYPE.
A) Enter set-up mode by: pressing CALIBRATE and LACQUER
NUMBER together. (PASS should appear in the large display,
if not press END and try set-up mode again) then enter
PASSWORD. (9963 if unchanged)
B) Press LACQUER NUMBER. (?? will appear in the small
display).
C) Enter new lacquer number by using NUMBER KEYS.
D) Press ENTER. (???? will appear on large display).
E) Enter set piece value: by using NUMBER KEYS.
F) Press ENTER.
G) Press END.
3. TO CALIBRATE A LACQUER TYPE.
A) Press CALIBRATE.
B) Press LACQUER NUMBER. (?? will appear in the small
display).
C) Enter new lacquer number by using NUMBER KEYS.
D) Press ENTER.
E) Place probe on known test sample. (remember to apply
pressure if the flat sheet probe is being used). Press ENTER
before releasing.
F) Press END.
4. TO TAKE READINGS.
A) Select appropriate lacquer number by using the UP/DOWN
ARROW KEY.
B) Place probe on sample.
C) Note readings. (Press ENTER if using printer or host
computer).