19
6 Troubleshooting
Display
Possible Cause
Suggestions
Please wait 30s
for sensor
stabilization!
During pH EasyCal or
Standard/Slope calibration,
when “Place Sensor in pH
Buffer #1(or 2)” is prompted,
the “Enter key is pressed
within 30 seconds.
Wait 30 seconds
Please Check
Sensor!
1. During pH EasyCal, the
detected sensor input is
more than 2PH away from
the selected pH value or
the range of possible
buffer values.
2. During pH Standard/Slope
calibration, the detected
sensor input is more than
2PH away from the
entered pH value.
3. During ORP One Point
calibration, the detected
sensor input is more than
120mV away from the
entered ORP value.
1. Check the sensor and restart
the calibration procedure.
2. Press “Enterto accept the
value.
Please Maintain
or Replace
Sensor!
1. During pH EasyCal, the
detected sensor input is
more than 4PH away from
the selected pH value or
the range of possible
buffer value.
2. During pH Standard/Slope
calibration, the detected
sensor input is more than
4PH away from the
entered pH value.
3. During ORP One Point
calibration, the detected
sensor input is more than
240mV away from the
entered ORP value.
1. Press “Enterto interrupt the
calibration. Replace the
sensor and restart the
calibration procedure.
2. Press “ESC” to reenter or
reselect the pH or ORP value.