The calibration looks like any
other scanning, the progress bar
showing the advancement of the
action.
The result of the calibration control is shown in a pop-up message and can be as following:
§
Calibration check passed
No changes are inferred.
§
Processing failed. No match to the polystyrene was obtained
Make sure that the Calibration cap is in position and repeat the calibration procedure.
Recurrent failure indicates instrument error.
Calibration Check Passed.
The instrument is ready to be
used.
Calibration Check Failed.
The obtained values for the poly-
styrene spectrum are not within
the acceptance criteria. Make
sure you reposition the calibration
cap and restart the process.