EasyManuals Logo

Servomex 2700 User Manual

Default Icon
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
Servomex 2700 Quickstart Manual
Servomex 2700 Quickstart Manual
3
2. STARTUP AND SHUTDOWN
2.1 Analyser startup
If the sensor head and sample probe have been removed from
the flue, then refit the sensor head and sample probe to the flue
and reconnect any interconnecting wiring (see Installation
Manual).
Switch on the electrical power to the sensor head and to the
control unit.
Leave the unit for two hours to warm up and then check that no
faults are indicated on the control unit display screen.
Connect the aspirator air supply to the sensor head. The correct
operating pressure is stated on a label attached to the inside of
the terminal enclosure on the side of the sensor head.
If necessary, connect calibration gases to the sensor head to
verify correct operation. See Section 3.4 for Calibration
procedure.
CAUTION
If the analyser is to be installed in a low temperature
environment (-20°C) then it should be powered up for a
period of two hours before the compressed air aspirator
supply is applied. This prevents corrosive sample gases
from condensing inside the sensor head pipe work.
NOTES
The analyser should be allowed to stabilise for several
hours (typically overnight) before calibration.
It is then recommended that calibration gases be passed
through the sensor head to confirm operation of the
analyser.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Servomex 2700 and is the answer not in the manual?

Servomex 2700 Specifications

General IconGeneral
TechnologyParamagnetic
Measurement Range0-100% O2
Outputs4-20 mA, RS232/RS485
Power Supply24 VDC or 100-240 VAC
Operating Temperature0 to 50°C (32 to 122°F)

Related product manuals