Servomex 2700 Quickstart Manual
Servomex 2700 Quickstart Manual
19
The software filter will also progressively reduce the output
fluctuation of the instrument as the filter setting is increased
from 0 to 9.
1. Press the MENU key, then select CONFIGURE, enter a
valid supervisor password, select SETUP.
2. From the SETUP menu, select FILTER.
3. If two sensors are fitted in the sensor head then select
OXYGEN or COe.
4. Input a value from 0 to 9, then press the QUIT or
MEASURE key when all is satisfactory.
The user may need to make several adjustments to optimise the
filter settings for their process conditions and control system
requirements.
3.7 Language options (Refer to Figure 3b)
The user is able to configure the display of the analyser to one
of three languages; English, French or German.
1. Press the MENU key, then select CONFIGURE, enter a
valid supervisor password and select SETUP.
2. From the SETUP menu, select UTILITY, then LANG and
then select ENGLISH, FRANCAIS, or DEUTSCH.
3. Press the QUIT or MEASURE key, when all is satisfactory.