n Inputting the tube volumeINNER VOLUME
Inputting a value in a low-pressure
gradient system enables the system
controller to calculate the system volume.
Refer to the table to input an appropriate
value according to the tubes used.
Input "0" in any mode other than the low-pressure gradient mode.
Tubing Between Solenoid Valve
and Pump Inlet
Tubing Between Pump Outlet
and Mixer Inlet
Set Value
Standard tube
(OUT tube, 310 mm)
SUS Tubing (ID 0.3 x 600) 200 µL
SUS Tubing (ID 0.3 x 300) 180 µL
Tube for buffer solution
(OUT tube, 400 mm)
SUS Tubing (ID 0.3 x 600) 355 µL
SUS Tubing (ID 0.3 x 300) 335 µL
Use the Mixer installation kit
(228-65020-41)
SUS Tubing (ID 0.3 x 185) 0 µL
The values above are intended when using the mixer recognition device. When the mixer
recognition device is not used, add the capacity depending on the mixer.
The column oven automatically recognise the capacity of the mixer when using the mixer
recognition device.
The values above are intended for the recommended tubes. When using a tube not listed
above, increase or decrease the capacity depending on the tube.
Example
1) Use the standard tube + SUS Tubing (ID 0.3 x 600) + Mixer MR40 LPGE (without the
mixer recognition device)
  →Set value: 240 (200 µL from the table above + 40 µL (mixer capacity))
2) Using the tube for buffer solution + SUS Tubing (ID 0.3 x 600) + Mixer MR40 LPGE
(with the mixer recognition device)
  →Set value: 355from the table above (The capacity of the mixer will be
automatically recognised by the column oven)
3) Use the mixer instllation kit (without the mixer recognition device)
  →Set value: 40 (0 µL from the table above + 40 µL (mixer capacity))
LC-40D X3 CL LC-40D XR CL
43
3.5 Auxiliary Functions Screen (FUNCTION)
3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu LC-40D X3 CL and is the answer not in the manual?

Shimadzu LC-40D X3 CL Specifications

General IconGeneral
BrandShimadzu
ModelLC-40D X3 CL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals