Avoid installing the instrument in dusty or corrosive gas locations to prolong service life.
Ensures proper ventilation, avoids open flames, and includes fire safety measures.
Advises against installing near strong magnetic equipment and suggests noise blocking.
Specifies firm, stable, flat surface requirements and clearance for maintenance.
Details temperature, humidity, overvoltage, pollution degree, and environmental factors.
Provides dimensions and weight for LCMS-8050/8045 and rotary pump.
Illustrates system layout and space requirements for top and front views.
Details space requirements for Micro LC unit (LC-Mikros, CTO-Mikros) with LCMS 8045/8050.
Guides on using the specified Shimadzu rotary pump for optimal performance and safety.
Instructs on connecting the PC using USB cables, with maximum length limits.
Specifies voltage, current, power capacity, and grounding requirements for the instrument.
Details the types of gases (Nitrogen, Dry Air, Argon) and their purity/pressure requirements.
Explains tubing preparation and connection for Nebulizer, Drying, and Heating gases.
Provides specifications for the PEAK Scientific GENIUS 1051 nitrogen/dry air generator.
Discusses using gas from centralized piping and the need for specific tubing.
Details connection adapters, regulators, and piping for gas cylinders.
Specifies tubing joint for existing nitrogen gas generators with 1/4" tubing.
Covers piping and connection for Argon gas used as CID gas.
Lists general safety precautions for handling high-pressure gas cylinders.
Details venting requirements for rotary pump and solvent vapors to exhaust facilities.
Lists standard samples, sensitivity check samples, and mobile phase components required.
Specifies drain bottles and reservoir bottles needed for operation.
Provides packing box dimensions, weight, and considerations for transport and handling.
Discusses sensitivity test criteria when installing to existing HPLC or LCMS systems.
Provides a checklist of installation conditions for the customer to verify before installation.