Installation
GMS815P-PS-2G/-3G · Supplementary Operating Instructions · 8016234 V2.0 · © SICK AG 15
Subject to change without notice
3.2 Assembly
3.2.1 Ensuring ambient conditions
Quiet running
▸
Protect the device against heavy jolts and vibrations (limit values → p. 33, §6.4).
Temperature
▸
Avoid enclosure exposure to direct sunlight.
▸
Maintain the allowable ambient temperature during operation (→ p. 33, §6.4).
Humidity
▸
Choose a dry installation location free from frost.
▸
Prevent moisture condensation – inside the device as well.
▸
Maintain the allowable relative air humidity (→ p. 33, §6.4).
Corrosive atmospheres
When the atmosphere at the installation location can contain corrosive gases:
▸
Install the Enclosure GMS815P-PS-2G/-3G in an outer housing (e.g. closed cabinet).
Purge the outer housing with a protective gas.
3.2.2 Fitting the enclosure
▸
Secure the enclosure on a structure that can safely carry the weight of the enclosure.
▸
Fit the enclosure so that the underside of the gas analyzer enclosure is more or less
horizontal (allowable offset → p. 33, §6.4).
3.2.3 Removing transport safety devices
1 Open the upper section of the gas analyzer enclosure (→ p. 22, §3.8.1).
2 Take the foam inserts out of the enclosure.
▸
Recommendation: Keep the foam inserts and use these every time the device is
transported.
3 Close the upper section of the gas analyzer enclosure again carefully; pay attention to
leak tightness (enclosure seal, screws on front door).
● Dimensions → p. 32, §6.1
● Weight → p. 33, §6.3