6 GMS815P-PS-2G/-3G · Supplementary Operating Instructions · 8016234 V2.0 · © SICK AG
Important Information
Subject to change without notice
1.1 Main hazards
Health risks through dangerous sample gases
GMS800 series” Operating Instructions
When shutting down
1.2 Main operating information
Before start-up
Make sure the internal mains switch is switched on ( p. 22, Fig. 4).
Close the enclosure tight.
If the enclosure is deformed or damaged: Do not put the GMS800 into operation and
secure against unauthorized start-up.
Start-up/operation
Carry out start-up and shutdown in accordance with the Operating Instructions deliv-
ered with the pressurized enclosure system.
Do not open the enclosure during operation.
Before shutting down: Purge the sample gas path with a dry neutral gas to prevent con-
densation in the measuring system.
Service information: Always wait at least 30 seconds between single switch-on actions
(cooling time for the switch-on current limiter).
In hazardous situations
Switch-off the emergency switch or main switch of the host system.
If liquid has penetrated the enclosure: Shut the GMS800 down immediately and dis-
connect mains voltage at external source.
WARNING: Risk of explosions
Do not open the enclosure when an explosive atmosphere is present.
Wait at least 60 minutes after disconnecting from the mains voltage before
opening the enclosure.
Observe the safety information on the enclosure.
The GMS800 is switched off automatically should the protective gas feed
fail.
Pressurized enclosure parameters p. 35, §6.8