5  Transducer Accessories and Biopsy 
5 - 10  Instructions for Use 
Biopsy (Puncture) Guideline Function 
  WARNING: Percutaneous procedures always involve heightened risk to the patient and to the 
operator handling biopsy needle guides. Clinicians using Siemens recommended biopsy devices 
under ultrasound guidance should be trained and must observe proper needle insertion 
sequencing with the needle guide in order to avoid undue discomfort and unnecessary risk and 
injury to the patient. 
  WARNING: The system displays biopsy guidelines to indicate an anticipated needle path. The 
guidelines do not account for a bent or deflected needle and are not intended as an absolute 
reference. It is the user's responsibility to verify correct positioning of the needle during a biopsy 
or puncture procedure. 
You can display on-screen guidelines for use with transducers compatible with needle guide 
attachments. 
Activating the On-screen Guidelines 
The biopsy function is activated only during real-time imaging in the following modes: 
  2D/Doppler    2D/Doppler with color 
  2D/M-mode    2D/Doppler with power 
  2D-mode with color    2D/Clarify 
Should you attempt to activate the biopsy function from an incompatible mode, the system 
displays a message to indicate which corrective action the system will take to allow you to 
activate the biopsy function.  
  Select OK to enter the biopsy function. 
  Select Cancel to leave the system as it is. The system will not display the biopsy 
guidelines on the screen. 
To display guidelines on the image screen: 
1.  Activate the biopsy function. 
●  Select Biopsy and then select A or B.  
○  For systems without a touch screen, select Biopsy A or Biopsy B on the image 
menu.  
Note: The selection corresponds to the track on the needle guide attachment. For fixed-angle 
needle guides, always use the Biopsy A selection. 
The system displays the following message as a precaution:  
Please verify that the physical needle guide matches your angle selection.  
2.  Ensure the selected angle (Biopsy A or Biopsy B) corresponds to the needle guide 
attached to the transducer and then click OK. 
3.  To remove the biopsy guidelines from the screen: 
●  Select the active biopsy and then select Off.  
○  For systems without a touch screen, select Biopsy Off on the image menu.