6 Transesophageal Transducer
Instructions for Use 6 - 37
V5Ms Transducer Technical Description
Transducer Type: Phased Array, multiplane
Elements: 64
Array Rotation Angle:
0° to 180°
Distal Tip Articulation: 120° anterior, 90° posterior
45° right, 45° left
Maximum insertion length:
1080 mm
Maximum insertion width:
15 mm
2D-mode Transmit
Frequencies:
4.0 MHz, 5.0 MHz, 6.0 MHz, 7.0 MHz
Color mode Transmit
Frequencies, including
CDV, Color DTI:
3.5 MHz, 3.75 MHz, 4.0 MHz
Doppler Transmit
Frequencies, including PW,
SCW, Aux CW,
Pulsed Wave DTI: 3.5 MHz, 3.75 MHz, 4.0 MHz
Imaging Modes:
2D-mode, PW Doppler, Steerable CW Doppler, Color mode, M-mode, Color DTI,
Pulsed Wave DTI, Anatomical M-mode
Maximum Image
Field of View: 90°
Image:
Sector
Operating Environment
Temperature:
Relative Humidity:
10°C to +40°C
10% to 80%
Storage Environment
Temperature:
Relative Humidity:
-10°C to +50°C
10% to 95%
Degree of protection
against electrical shock:
V5Ms transducers with this label are certified as type BF
defibrillator proof per IEC 60601-1 Standard for Safety of
Medical Equipment
V5Ms transducers with this label are certified as type BF (not
defibrillator proof) per IEC 60601-1 Standard for Safety of
Medical Equipment
Degree of protection
against ingress of fluid: IPX8

Table of Contents