2 Safety and Care
Instructions for Use 2 - 19
Transducer Surface Temperature Limits
The following table provides the maximum surface temperature of the transducers compatible
with the system.
Maximum surface temperatures are in accordance with IEC 60601-2-37.
Maximum Temperature
Transducer
TMM
Still Air
4C1
38.6°C
35.9°C
6C1 HD
41.0°C 38.0°C
6C2
40.1°C 40.0°C
8C3 HD
41.0°C 42.5°C
9L4
39.4°C 38.1°C
12L4
42.5°C 47.0°C
14L5
41.7°C 41.1°C
14L5 SP
40.7°C
40.5°C
18L6 HD
42.9°C 43.0°C
4P1
40.9°C 42.6°C
4V1
40.3°C 40.8°C
4V1c
42.3°C 46.0°C
8V3
41.4°C
41.0°C
10V4
37.7°C 36.5°C
V5Ms
36.4°C 39.4°C
V7M
42.6°C 40.3°C
EV-8C4
38.9°C 39.6°C
EC9-4
36.0°C
36.7°C
MC9-4
42.5°C 45.3°C
7CF1
38.6°C 45.0°C
7CF2
36.7°C 37.2°C
9EVF4
38.4°C 38.3°C
CW2
36.9°C 37.7°C
CW5
35.3°C 37.7°C
AcuNav 8F
41.1°C
NA
AcuNav 10F
39.6°C
NA
TMM = Tissue Mimicking Material
NA = not applicable (not tested)