2  Safety and Care 
Instructions for Use  2 - 49 
Cleaning, Disinfecting, and Sterilizing Transducer Accessories 
  WARNING: Ensure the accessories for a transducer are properly cleaned, sterilized, or 
disinfected as appropriate before each use to avoid possible patient contamination. 
Needle Guide Bracket Kits 
Needle guide bracket kits are available for specific transducers. Instructions follow for the 
cleaning, disinfecting, and sterilization of each kit. Bracket assemblies should be cleaned and 
sterilized or high-level disinfected after each use. 
See also: For a list of transducers compatible with the needle guide accessories, refer to Chapter 5 
in this manual. 
EC-1 Needle Guide Bracket Kit for Endocavity Transducers 
  WARNING: The EC-1 Needle Guide Bracket kit is packaged non-sterile. Sterilize this product 
prior to its first use. 
Prior to sterilization, clean the bracket assembly. 
To clean the bracket assembly: 
1.  Rinse the bracket under water and remove all debris. Avoid using a brush as it may 
damage the needle guide bracket. 
2.  Visually inspect the guide to make sure all debris is removed. 
3.  Wipe off water on the bracket assembly using a sterile cloth or sterile gauze. 
To sterilize the bracket assembly: 
Use sterile technique to perform this procedure. 
1.  Sterilize assembly using high-pressure steam sterilization (130°C [266°F] for an exposure 
time of 10 minutes), EO (ethylene oxide) sterilization, or hydrogen peroxide plasma 
sterilization. 
2.  Perform one of the gas sterilization methods listed below: 
 
Sterilization 
Gas 
Ethylene Oxide Gas  10%  50°C (122°F)  50% 
Maximum operating 
pressure 980 hPa 
2
 
7 hours 
Hydrogen Peroxide 
Plasma 
STERRAD 
For details, refer to the operation manual for the STERRAD
 
sterilization system. 
3.  After gas sterilization, thoroughly degas (aerate) the biopsy adapter to remove all gas 
residues.