Appendix E Advanced Feature Controls
Instructions for Use E - 17
Selection
Description
CPS Capture (Available only for systems without a touch screen)
Starts or stops the acquisition for the contrast pulse sequence (CPS) or contrast harmonic
imaging (CHI) technique.
Duration Rotate:
Selects the duration of CPS acquisition in seconds (sec).
1, 2, 3, 5, 10, 15, 20, indefinite
Note: For systems without a touch screen, the selection is
assigned to the scroll wheel.
Press:
Starts or stops the acquisition for the contrast pulse sequence
(CPS) or contrast harmonic imaging (CHI) technique.
Power Adjusts the transmit power during contrast imaging.
0.10% to 100% dependent on exam type
Note: For systems without a touch screen, the selection is on the LED panel.
See also: Transmit Power Control, Safety and Care, Chapter 2, Instructions for Use
Protocol Selects the clip capture duration in seconds (sec).
30, 45, 60, 90, user-defined
ADI Enables the Agent Detection Imaging (ADI) technique during contrast agent imaging.
On
Off
CPS/CHI (Available only when ADI is disabled)
Enables the contrast pulse sequence (CPS) or contrast harmonic imaging (CHI) technique.
CPS
CHI
Note: For systems without a touch screen, the selection is on the LED panel.
2D/CA/Mix
Displays the 2D image (2D), the contrast agent image (CA), or a combination (Mix) of the
contrast agent image with the 2D image.
Note:
For systems without a touch screen, the selection is on the LED panel.
Live Dual Live Dual
Enables or disables the display of two real-time images originating from the same acquisition in
a side-by-side format.
On
Off