12 Adjustments
ARCADIS Varic / Orbic SPR2-310.842.01.07.02 Siemens AG
12.06 CS SD 24
Page 12 of 34
Medical Solutions
7. Click on <Go>.
¹ Status bar: Init Off & Store Positions --> running --> success A window appears:
Calibration is finished --> confirm with <OK>.
¹ The main system restarts.
8. Manually set the image rotation to the 0 degree position at the control console.
¹ See display for image rotation at the control console.
X 9. Briefly release fluoro.
¹ The image should now display in the correct 0 degree position (Fig.4/p.11); if
not, repeat adjustment.
10. Click on “>” for the next page and click on <Finish>.