Siemens AG SPR2-310.842.01.07.02 ARCADIS Varic / Orbic
12.06 CS SD 24
Adjustments 15
Page 15 of 34
Medical Solutions
Display X-iris 2.5
• Use the X-iris keys at the control console for the adjustment.
• “Collimator X-Iris” has been adjusted correctly.
1. Register an emergency patient to allow radiation release.
2. Click on <Go>.
¹ Status bar: Init --> running --> success
X 3. Briefly release fluoro for the LIH image.
4. Adjust the graphic display until the graphic circle is aligned with the circle size of the
X-iris diaphragm.
5. Click on <Go>.
¹ Status bar: Open Pos-. I.I. - Full Format --> running --> success
X 6. Briefly release fluoro for the LIH image.
7. Adjust the graphic display until the graphic circle is aligned with the circle size of the
X-iris diaphragm.
8. Click on <Go>.
¹ Status bar: Close Pos. I.I. - Full Format --> running --> success
X 9. Briefly release fluoro for the LIH image.
10. Adjust the graphic display until the graphic circle is aligned with the circle size of the
X-iris diaphragm.
11. Click on <Go>
¹ Status bar: Close Pos. I.I. - Zoom Format --> running --> success
X 12. Briefly release fluoro for the LIH image
13. Adjust the graphic display until the graphic circle is aligned with the circle size of the
X-iris diaphragm.
14. Click on <Go>.
¹ Status bar: Open Pos. I.I. - Zoom Format --> running --> success
Prerequisites
Adjustment
Open Pos. I.I. Full Format
Close Pos. I.I. Full Format
Close Pos. I.I. Zoom Format
Open Pos. I.I. Zoom Format