18 Adjustments
ARCADIS Varic / Orbic SPR2-310.842.01.07.02 Siemens AG
12.06 CS SD 24
Page 18 of 34
Medical Solutions
9. Click on <Go>.
¹ Status bar: Init Off & Store Position --> running --> success
¹ A window appears: Calibration is finished --> confirm with <OK>.
¹ The main system restarts.
10. Click on “>” for the next page and click on <Finish>.
11. Check the adjustment and repeat if necessary.