Siemens AG SPR2-310.842.01.07.02 ARCADIS Varic / Orbic
12.06 CS SD 24
Adjustments 19
Page 19 of 34
Medical Solutions
Display Slot Diaphragm 2.7
• “Collimator Slot Diaphragm” has been adjusted correctly.
• Use the slot diaphragm keys at the control console for the adjustment.
• Use 2.1 mm Cu as prefilter.
1. Register an emergency patient to allow radiation release.
2. Click on <Go>.
¹ Status bar: Init --> running --> success
¹ Press the collimator X-iris “open button” at the control console to open the X-iris
completely.
X 3. Briefly release fluoro for the LIH image.
4. Adjust the display lines on the monitor until the lines are aligned with the front edges of
the slot diaphragm blades.
5. Click on <Go>.
¹ Status bar: Open Pos. I.I. - Full Format Position --> running --> success
X 6. Briefly release fluoro for the LIH image.
7. Adjust the display lines on the monitor until the lines are aligned with the front edges of
the slot diaphragm blades.
8. Click on <Go>.
¹ Status bar: Close Pos. I.I. - Full Format --> running --> success
X 9. Briefly release fluoro for the LIH image.
10. Adjust the display lines on the monitor until the lines are aligned with the front edges of
the slot diaphragm blades.
11. Click on <Go>.
¹ Status bar: Close Pos. I.I. - Zoom Format --> running --> success
Prerequisites
Adjustment
Open Pos. I.I. Full Format
Close Pos. I.I. Full Format
Close Pos. I.I. Zoom Format
Open Pos. I.I. Zoom Format