Hematek
®
3000 System – Instruction Manual – Version 1.3 33
1. Make sure that the electrical and environmental conditions are met, see
Electrical on page 9-3 and Environmental on page 9-3.
2. Make sure that the inlet area for the blower type fan underneath the
system is not blocked.
Remove transportation locks 3. At the front of the system, remove the adhesive tape securing the slide
drawer and waste tank.
4. At the top of the system, remove the two adhesive tapes securing the lid.
5. Open the lid.
6. In the pocket at the rear of the slide stainer, remove the knurled thumb
screw together with the red cardboard strip.
7. Remove the middle cover from the system.
8. Remove the thin foam sheet from the system.
6 Platen
7 Leveling feet
8 Sensing switch fingers
9 Slide drawer
10 Waste tank
Figure 3-1 Installing the system (continued)