29
Siemens AG Healthcare Sector 11/12 Technical data
Dimensions
Component
Width
[cm]
Depth
[cm]
Height
[cm]
Weight
[kg]
Heat
dissipa-
tion
[kW]
Examination Room
Magnet 3 Tesla AS
(incl. Helium)
205 163 215 5755
Magnet in operation,
incl. gradient coil,
body coil, Tim table,
and covers
231 433
461
1
1. With Tim Whole Body Suite option
219 7320
Tim table 76 249 52–102
2
+13 mm
3
2. Including Heightening Kit, if necessary
3. Depending on the floor conditions
Required min. room
height clearance
240
4
4. Finished floor to finished ceiling
Min. transport dimen-
sions
231 182 227
Control Room
syngo Acquisition
Workplace
(table+monitor)
12080117
(72+45)
Host computer 22 46 47
syngo MR Workplace
(optional)
(table+monitor)
12080117
(72+45)
Equipment Room
Electronics cabinet,
incl. system control,
RF system, gradient
power system, image
processor
160 65 198
5
5. Without attachments
1500 5
6
6. Only ventilation might be required
Cooling system 65 65 189 500