Siemens AG Healthcare Sector 11/12 Technical data
Line power supply
For all products, line power has to be supplied via an on-site system contact or via another
multipole shut-down mechanism. Room installation has to be in compliance with
VDE 0100-710
1
.
1. In all countries, compliance with local and national legal regulations is required.
However, we strongly recommend compliance with the regulations described herein – to the extent per-
mitted by relevant local and national laws – in order to ensure the safety of operating personnel, patients,
and third parties.
Power rating plate
XQ gradients Values Tolerance
Voltage 380 V, 400 V, 420 V, 440 V,
460 V, 480 V
±10%
Frequency 50/60 Hz ±1 Hz
Connection value 110 kVA
On-site protection 160 A