Siemens AG Healthcare Sector 11/12 Technical data
26
Specific absorption rates
The specific absorption rate for the RF power and the rate of change of the gradient fields
are checked according to the requirements of norm IEC 60601-2-33.
Cooling system
Two different customer specific cooling alternatives (Separator or Eco Chiller) are avail-
able.
Separator option (for connec-
tion to available cooling sys-
tem)
Water consumption 90 l/min
1
1. Water temperature: 12 °C (45 °F)
Heat dissipation to water 60 kW
Eco Chiller option with auto-
matic adaptation to the
required cooling demands (e.g.
different night/day mode) to
decrease energy cost
GREEN Cooling Package
2
:
• Automatic start if the surrounding temperature is
18 °C (64 °F) or less
• If the temperature is less than –10 °C (14 °F) the
chiller is switched off
3
2. Free Cooling Unit, optional
3. In case of clinical routine measurement conditions