Operator Responsibilities for Maintenance 5
© Siemens, 2009 M7-000.664.01.04.02 Page 5 of 14 MAGNETOM
03.12 CS SD MR
Safety check (including safety-related tests)
The following checks are intended to ensure the safety of the system. Where appropriate,
preventive measures have to be adopted or repairs performed. The points to be checked
are generally regulated by laws and standards.
The Safety-related Tests according to §6 MPBetreibV (Germany) are mandatory.
The specified checks and intervals correspond to the minimum requirements. Compliance
with stricter national legislation may occasionally be necessary.
Object or function Reason What is checked: Interval
General tests
System Safety of patient, personnel, and
system
Visual inspection of the system Annually
Visible cabling, cable
routing
Safety of patient, personnel, and
system
Visual inspection of cables,
cable routing
Annually
Options Safety of patient, personnel, and
system
Visual inspection of options Annually
Accessories Safety of patient, personnel, and
system
Visual inspection of accessories
1
Annually
Magnet Optimum function within speci-
fications
Safety of patient, personnel, and
system
Check pressure of the magnet
Check for ice formation
Check for leaks
Annually