Operator Responsibilities for Maintenance 11
© Siemens, 2009 M7-000.664.01.04.02 Page 11 of 14 MAGNETOM
03.12 CS SD MR
Preventive maintenance
The purpose of preventive maintenance is to keep unforeseen failures to a minimum. This
satisfies the prerequisites for system compliance with the guaranteed, long-term charac-
teristics.
The effects of different operating conditions (full or partial load operation, temperature,
size of dust particles, humidity, gases, vapors) are checked and the condition of parts sub-
ject to wear is determined by recording and analyzing characteristic values. Preventive
measures must be adopted or repairs must be undertaken as appropriate.
The specified maintenance intervals correspond to the minimum requirements. Compli-
ance with stricter national legislation may occasionally be necessary.
Object or function Reason What is checked: Interval
Cooling system Preventive measure to
avoid overheating
Check for leaks and condensation
Check/replace water filter
Check water pressure and add water if
necessary
Annually
Air filters, fans Preventive measure to
avoid pollution
Replace air filter
Check functionality of fans
Annually
Chiller (option) Preventive measure to
avoid overheating
11