Loading...
MAGNETOM M7-000.664.01.04.02 Page 8 of 14 © Siemens, 2009
03.12 CS SD MR
8 Operator Responsibilities for Maintenance
Tim dockable table
(option)
Safety of patient and personnel Inspect table movement and
end switches
Operation of the emergency
stop and safety switches
Operation of the emergency
release
Checking tolerance limits
Marking of the hazard area
Operation of the locking mech-
anism
Inspect the emergency undock-
ing function
Checking the chassis and the
wheels
Annually
Intercom, pneumatic
bulb
Communication with the patient Function of intercom
Function of pneumatic bulb
Annually
Monitoring the magnet
(MSUP)
Magnet stop function
(ERDU)
Prevent hazards caused by mag-
net
Operation of the magnet super-
vision and the magnet stop
Annually
Gradient supervision Optimum function within speci-
fications
Safety of patient, personnel, and
system
Function of the gradient super-
vision
Annually
QA (Quality Assurance)
measurements
Optimum function within speci-
fications
Patient protection
Interaction of various system
components according to the
guaranteed system characteris-
tics using the complete QA mea-
surement
Annually
1. According to manufacturer recommendations
2. If installed
Object or function Reason What is checked: Interval