Safety information for the system owner
0.0
A.1-15
0.0
syngo MR D13
0.0
A.1
Examination room and ambient
conditions
A.1
Explosion protection
A.1
The MR system is not intended for operation in areas prone to
explosive anesthetic gases.
A.1
Emergency switches
A.1
The voltage to the MR system can be turned off via a Power-Off
(System Off or Emergency Shut-down) switch installed
on-site. The switch can be used to stop the scan immediately in
case of emergency. The room installation must correspond to
VDE 0100-710 and/or national laws.
A.1
A quench can be released with the Magnet-Stop switch. This
switch is installed on-site as well.
A.1
Air conditioning
A.1
An air conditioning system must be used to ensure the required
environmental conditions. (Technical data)
A.1
A.1
The functions of the air conditioning or the temperature and rel-
ative humidity of the examination room must be checked at reg-
ular intervals.
A.1
The air conditioning is installed on-site by the system owner.
It is not part of the MR system. Information with respect to
maintenance (e.g. replacing filters) and monitoring the
functions of the air conditioning are included in the operat-
ing instructions of the air conditioning manufacturer.