Siemens AG Healthcare Sector 11/12 MR compatibility data sheet
10
In the following we show a plot representing the 0.5 T, 1 T, 1.5 T, and 2 T iso-magnetic
contours at positions accessible to and relevant for the MR worker as far as the static mag-
netic field in the isocenter exceeds any of these values.
⊗: At this location, the value of the magnetic field B
o
is greatest.
! The iso-magnetic contour lines are accurate to a value better than 1%. There is very lit-
tle influence from the environment to the scanner. The graphic shows the cover in
nominal position. On a variety of scanners, the position can vary ±5 mm in the axial
direction. The maximum value therefore has a tolerance of ±10%, as small geometric
deviations can cause a significant change in the value.
−
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2
1.2
1.4
1.6
1.8
0.2
0
0.2
0.4
0.6
0.8
1
1.5 T
1 T
0.5 T
2 T
3.4 T
Z Axis (m)
Vertical (m)
Patient table
Side view