Installation and setting 2 - 5
Siemens-Elema AB SPR8-220.033.03 Page 5 of 12 MOBILETT Plus HP
Solna, Sweden Rev. 02 10.02 SPS-UD
Capacitor formation (if necessary) 2
Remove the system upper cover
Switch "1" of S1 on D1 to "ON".
w This will display the program version:
kVmAs
vErX.X = Version X.X (depends on software version installed).
Select test programs by pressing "kV+" button until "P06" is displayed.
Press lamp button to start the capacitor formation.
w The test program starts with displaying:
kVmAs
CAPFor = Capacitor formation program
w After a short while, starts the program to display the actual voltage in the
capacitor bank:
kVmAs
CAP150 = 150 Volt.
Approximately every 6 minute the voltage values is increased. The charging sequence is
increased as follows:
w 150 - 250 - 299 - 325 - 337 - 345 - 348 - 351 - 353
At the 353 V level the charging will continue for another 48 minutes to decrease and stabi-
lize the capacitors internal leakage current.
w If the formation is successfully finished the display will show:
kVmAs
P06 PASS = Capacitor formation completed.
Duration: Approximately 100 minutes.
Switch "1" of S1 on D1 to "OFF"

Other manuals for Siemens MOBILETT Plus HP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens MOBILETT Plus HP and is the answer not in the manual?

Siemens MOBILETT Plus HP Specifications

General IconGeneral
BrandSiemens
ModelMOBILETT Plus HP
CategoryMedical Equipment
LanguageEnglish

Related product manuals