General 1 - 3
Siemens-Elema AB SPR8-220.033.03 Page 3 of 8 MOBILETT Plus HP
Solna, Sweden Rev. 02 10.02 SPS-UD
Safety information and protective measures 1
• When performing service work and tests adhere to:
- the product-specific safety information in the documents,
- as well as the general safety information contained in ARTD
Part 2.
• Connect the MOBILETT Plus HP only to mains power supply
outlet (receptacle), that corresponds to the installation
requirements of VDE 0107 or country-specific regulations.
• Remove or install boards only when the generator is switched
OFF. Adhere to the ESD guidelines .
• Checks and adjustments performed with radiation ON are
identified by the radiation warning symbol . During these
types of adjustments, radiation protective clothing must be
worn.
CAUTION