4 - 1
Siemens-Elema AB SPR8-220.033.03 Page 1 of 2 MOBILETT Plus HP
Solna, Sweden Rev. 02 10.02 SPS-UD
Test certificate, MOBILETT Plus HP 4
This test certificate is subdivided by analogy to installation and adjustment instruction.
This test certificate shall be completed in duplicate, one copy remaining with the client, the
other to be filed at the branch office.
Part No./Type No./Serial No.: ............................................................................................
Site/System Designation: .....................................................................................................
Customer: .................................................. Customer No: .............................................
Unpacking, Warning labels
Labels
Battery indicator
Initialization selftest
Capacitor formation (Formation program used)
Functional tests with motor drive
kV and mAs indication, technique factors
Radiation Indicators
Manual termination of exposure
Remote exposure switch system (optional)
DAP measuring system (optional)
Test of kV-accuracy
Main switch Exposure values Measured kV
B 52 kV / 50 mAs .................................................
B 81 kV / 20 mAs .................................................
B 133 kV / 12.5 mAs .................................................
M 81 kV / 20 mAs .................................................
Measurement
device
Type: .................................
Serial No.: ..........................
Date calibrated: ..................
Test of mAs-accuracy
Main switch Exposure values Calculated mAs
B 40 kV / 5 mAs .................................................
B 81 kV / 2 mAs .................................................
B 133 kV / 10 mAs .................................................
M 81 kV / 2 mAs .................................................
Measurement
device
Type: .................................
Serial No.: ..........................
U
S
A