26 General Information on Troubleshooting
MOBILETT XP SPR8-230.840.01.05.02 Siemens AG
11.05 CS SD 24
Page 26 of 72
Medical Solutions
- Starting of the HV inverter
- Monitoring and controlling of the actual tube current
- Monitoring and controlling of the high-voltage
- Change after exposure into standby mode
- Anode braking
If the exposure is interrupted, the following error messages are possible:
- ERR 31 kV limitation
- ERR 34 tube current limitation
- ERR 35 short-circuit at the main inverter
- ERR 36 short-circuit at the filament inverter
- ERR 37 error in the kV control loop
- ERR 38 error in the mA control loop
- ERR 39 timeout "t-max exposure"
- ERR 41 exposure not completed
- ERR 42 filament current exceeds permissible limit