46 Function Check
MOBILETT XP SPR8-230.840.01.05.02 Siemens AG
11.05 CS SD 24
Page 46 of 72
Medical Solutions
6-
5Function Check
Reading out the MOBILETT XP error memory 0
• Switch off the system and disconnect the power plug if necessary. Set the XP Hybrid
operating mode switch to "0". Please observe all safety information.
• Remove the upper cover. Remove the XP Digital upper covers and place them on their
sides on a table or similar surface. Do not disconnect any cables.
• Move switch SW2B on CPU board D916 to position 2 (service) ((Fig. 15 / p. 46)).
• Switch on the system.
• Service program Pr 1 is displayed on the control panel. Use the mAs +/- keys to select
service program 3. Read out the error memory and make a note of the data (see the
service programs section of these instructions).
• Subsequently set switch SW2B of D916 to position 1 (user) (Fig. 15 / p. 46).
Fig. 15: CPU D916