34 Test Points, Fuses and LEDs
MOBILETT XP SPR8-230.840.01.05.02 Siemens AG
11.05 CS SD 24
Page 34 of 72
Medical Solutions
Fig. 13: D982 Battery charger__
D102 (motor control) 0
Potentiometer, fuse
V80+V40 Defective batteries in BK1 or BK2; or
fuse F3 blew
LED on
V34 Charger 2 is working; voltage present
at output (approx. 175 V max.)
LED on
P7 Output voltage (BK3 & BK4) Factory-only setting
Name Function / information Value
P1 Speed setting Factory-only setting
F1 Input 300 V (DC) 3.15 A fast