EasyManua.ls Logo

Siemens MOBILETT XP - Mode of Operation of the Capacitor Bank

Siemens MOBILETT XP
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Siemens AG SPR8-230.840.01.05.02 MOBILETT XP
11.05 CS SD 24
Basic functions 55
Page 55 of 72
Medical Solutions
Mode of operation of the capacitor bank 0
Capacitor bank D972 generates the intermediate circuit voltage for the HV inverter, fila-
ment inverter, and anode starter modules.
Capacitor bank D972 is continuously charged by capacitor charger D952.
Capacitor bank D972 consists of 12 capacitors connected in parallel that deliver sufficient
energy to operate the inverter. Each capacitor is connected in series with a fast fuse.
Each capacitor has a capacity of 10 mF. The total capacity is 0.12 F. To facilitate handling
of these live components, the LEDs at the capacitors show the charge level > 40 V.
When the system is "OFF", the capacitors are discharged via three resistors (R3, R4, R5)
within 15-20 minutes.
The charging time of the capacitors varies depending on the charge level of the capacitor
bank. Around 7 to 9 seconds are required for charging empty capacitors.
The intermediate circuit voltage of the capacitor bank is monitored by D916 (CPU).
In the case of an error, the charging procedure is interrupted.
When replacing one or more capacitors of the capacitor bank, all capacitors must have a
voltage value of < 2 V to avoid outage of the fast fuses on D972.

Table of Contents

Other manuals for Siemens MOBILETT XP

Related product manuals