70 Service programs
MOBILETT XP SPR8-230.840.01.05.02 Siemens AG
11.05 CS SD 24
Page 70 of 72
Medical Solutions
Program 7 - setting the maximum main inverter frequency 0
Program 8 - Anode starter - start/stop test 0
Program 9 - Filament current test 0
Program 10 - AutoOff on/off (Hybrid/Digital in battery mode) 0
Program 7
• After the start of the service program, "ADJF" is shown on the display.
• Measure F (max) at the test point TP-FC2 on D916.
• Set exposure switch S27 to "preparation".
• set the frequency of the main inverter with potentiometer P4 at D916.
(Turn counter-clockwise - Increase).
- Set value TP-FC2 is 50 kHz +/-0.3 kHz
Program 8
• "ROT220“ (rotation 220/sec.) is displayed on the display.
• Set exposure switch S27 to "preparation". The anode starter is activated. The anode
accelerates.
Program 9
• After the start of the service program, "FIL3“ is shown on the display.
• Set exposure switch S27 to "preparation". Preheating begins. "FIL5" is shown on the
display as long as exposure switch S27 is pressed.
Program 10
• After the start of the service program, "ON" or “OFF“ is shown on the display.
• ON = automatic switch-off after approx. 15 minutes of inactivity (default factory set-
ting: Hybrid = ON, Digital = OFF)
• Use the mAs+ or mAs- keys to switch over.
NOTE:
If the battery voltage is less than 110 V (right or left side), the system is always switched
off (Hybrid and Digital).