70 Service programs
MOBILETT XP SPR8-230.840.01.05.02 Siemens AG
11.05 CS SD 24
Page 70 of 72
Medical Solutions
Program 7 - setting the maximum main inverter frequency 0
Program 8 - Anode starter - start/stop test 0
Program 9 - Filament current test 0
Program 10 - AutoOff on/off (Hybrid/Digital in battery mode) 0
Program 7
After the start of the service program, "ADJF" is shown on the display.
Measure F (max) at the test point TP-FC2 on D916.
Set exposure switch S27 to "preparation".
set the frequency of the main inverter with potentiometer P4 at D916.
(Turn counter-clockwise - Increase).
- Set value TP-FC2 is 50 kHz +/-0.3 kHz
Program 8
"ROT220“ (rotation 220/sec.) is displayed on the display.
Set exposure switch S27 to "preparation". The anode starter is activated. The anode
accelerates.
Program 9
After the start of the service program, "FIL3“ is shown on the display.
Set exposure switch S27 to "preparation". Preheating begins. "FIL5" is shown on the
display as long as exposure switch S27 is pressed.
Program 10
After the start of the service program, "ON" or “OFF“ is shown on the display.
ON = automatic switch-off after approx. 15 minutes of inactivity (default factory set-
ting: Hybrid = ON, Digital = OFF)
Use the mAs+ or mAs- keys to switch over.
NOTE:
If the battery voltage is less than 110 V (right or left side), the system is always switched
off (Hybrid and Digital).