62 Basic functions
MOBILETT XP SPR8-230.840.01.05.02 Siemens AG
11.05 CS SD 24
Page 62 of 72
Medical Solutions
Sample heating measurements 0
Measuring the standby heating
Switch off the system
Remove the top cover and, in the case of MOBILETT XP Digital, place it on its side on a
table or similar surface (do not disconnect any cable connections).
Oscilloscope channel 1 at D916 TP “I” (7) and D916.GND (1) - resonant filament cur-
rent.
Oscilloscope setting: channel 1 = 2 V/div, trigger channel 1, trigger stage 1.5 V, 100
µs/div
Switch the system on
¹ Below is the oscillogram of the resonant filament current during standby
Fig. 25: Stand_by_operation_ih_ir
NOTE
The oscillogram below shows a single oscillation (shorter time
base) of the resonant filament current during standby.