Siemens AG SPR8-230.840.01.05.02 MOBILETT XP
11.05 CS SD 24
Basic functions 63
Page 63 of 72
Medical Solutions
Fig. 26: Stand By_I
Setting the highest frequency of the filament inverter (75 kHz)
The settings are made via a 2-channel oscilloscope and potentiometer P3 of D916 (CPU).
Switch off the system
Remove the top cover and, in the case of MOBILETT XP Digital, place it on its side on a
table or similar surface (do not disconnect any cable connections).
Remove the back cover of the MOBILETT XP (Access to D927)
Remove the metal plate over board D916 For MOBILETT XP Digital, remove the CXDI
PC (laptop) and also remove the holder plate
Oscilloscope channel 1 at D916 .FC1 (TP25) and D916.GND (TP1)
Oscilloscope setting: 1V/div, sweep 5 µs/div
Remove the protective cover of D927 Caution! It is possible to come into contact with
components under voltage!
Disconnect plug X5 of board D927 (breaking the filament circuit)
Move switch SW2B on D916 to 2 (service program) and switch the system on
Select service program 9
Use exposure switch S27 to press and hold "PREP"
Set the resonance frequency of the filament inverter via potentiometer P3 on D916 to
75kHz +/- 2 kHz