64 Basic functions
MOBILETT XP SPR8-230.840.01.05.02 Siemens AG
11.05 CS SD 24
Page 64 of 72
Medical Solutions
Fig. 27: Test point FC1 75 kHz
Measuring preheating
Switch off the system
Remove the top cover and, in the case of MOBILETT XP Digital, place it on its side on a
table or similar surface (do not disconnect any cable connections).
Oscilloscope channel 1 at D916 TP “I” (7) and D916.GND (1) - resonant filament cur-
rent
Oscilloscope setting: trigger channel 1, trigger stage 1 V, 2 V/div, sweep 100 µs/div
Select 81 kV and 5 mAs on the control panel In the case of XP Digital, switch to the
manual operating mode
Use exposure switch S27 to press and hold "PREP"
¹ Below is the oscillogram of the resonant filament current during preheating
NOTE
See the "MOBILETT XP, Replacement of parts" instructions for in-
formation on setting the filament current on the D916.