68 Service programs
MOBILETT XP SPR8-230.840.01.05.02 Siemens AG
11.05 CS SD 24
Page 68 of 72
Medical Solutions
8-
7Service programs
Overview 0
Activating service programs 0
System OFF
Remove the top cover.
Service switch SW2B on board D916 in position 2
System ON.
¹ Display shows "Pr 1"
Use the "mAs+ / mAs-" keys to select the service program
Use the Collimator Lamp button on the control panel to start a service program. To end
a service program, press the Collimator Lamp button again.
NOTE
If a service program is activated, no radiation can be released.
Program 1 - formatting the capacitor bank 0
Program 2 - Exposure counter 0
Program 1
After the start of the service program, "CAP 150" is shown on the display.
The voltage rises incrementally from 150 V to 440 V DC. Each program step is per-
formed for approx. 6 minutes.
The system maintains the voltage level of 440 V DC for approx. 1 hour. The program
is then ended automatically. After termination of the program, the message "CAP
END“ is displayed.
NOTE:
If the actual voltage is greater than 150 V, the program first performs a discharge to <
150 V. This process takes 10 to 15 minutes. The program then begins charging the
capacitor bank.
Program 2
Display of exposure number in the display
NOTE:
It is not possible to reset to zero in the event of a tube change.
The counter reading must be noted as the starting value for a new component (e.g. sin-
gle tank).