38 Error messages
MOBILETT XP SPR8-230.840.01.05.02 Siemens AG
11.05 CS SD 24
Page 38 of 72
Medical Solutions
ERR 9 Static kV measured value (tube current) <> 0
System
mode:
Standby
Explanation: A tube current is detected during stand-by. D916 (CPU)
detects actual mA error.
Action:
Check the cable connection between measurement data log D907 (the
single tank) and D916.X8 (CPU).
Possible causes: Cable connection between the single tank and D916 (X8);
"ground connection" from the system to the single tank is defective; D916;
D907; single tank.
ERR 10 Defect in a battery pack
System
mode:
Standby
Explanation: D916 (CPU) detects via D982 a battery pack with a voltage <=
125V or a defective battery pack.
Action:
Check the battery voltage, nominal value > 125 V, possibly charge bat-
tery.
Check fuse F5 on D982 (battery block output).
Check fuse F3 on D982 (charger 1).
Check fuse F4 on D982 (charger 2).
Check the cable connection between the battery blocks and D982.
Check the cable/contacts between the individual batteries.
ERR 11 Capacitor bank voltage too high
System
mode:
Standby
Explanation: D916 (CPU) detects voltage too high (> 460 V) on the capaci-
tor bank.
Action:
Check measurement point TP-VC on D916 (capacitor bank voltage)
Possible causes: D952
Rare causes: Cable connection X20 (D916-D952-D962); D916