38 Error messages
MOBILETT XP SPR8-230.840.01.05.02 Siemens AG
11.05 CS SD 24
Page 38 of 72
Medical Solutions
ERR 9 Static kV measured value (tube current) <> 0
System
mode:
Standby
Explanation: A tube current is detected during stand-by. D916 (CPU)
detects actual mA error.
Action:
• Check the cable connection between measurement data log D907 (the
single tank) and D916.X8 (CPU).
Possible causes: Cable connection between the single tank and D916 (X8);
"ground connection" from the system to the single tank is defective; D916;
D907; single tank.
ERR 10 Defect in a battery pack
System
mode:
Standby
Explanation: D916 (CPU) detects via D982 a battery pack with a voltage <=
125V or a defective battery pack.
Action:
• Check the battery voltage, nominal value > 125 V, possibly charge bat-
tery.
• Check fuse F5 on D982 (battery block output).
• Check fuse F3 on D982 (charger 1).
• Check fuse F4 on D982 (charger 2).
• Check the cable connection between the battery blocks and D982.
• Check the cable/contacts between the individual batteries.
ERR 11 Capacitor bank voltage too high
System
mode:
Standby
Explanation: D916 (CPU) detects voltage too high (> 460 V) on the capaci-
tor bank.
Action:
• Check measurement point TP-VC on D916 (capacitor bank voltage)
Possible causes: D952
Rare causes: Cable connection X20 (D916-D952-D962); D916