Siemens AG SPR8-230.840.01.05.02 MOBILETT XP
11.05 CS SD 24
Basic functions 59
Page 59 of 72
Medical Solutions
Fig. 23: Exposure_kVp_kVn_s
Setting the highest frequency for the main inverter (50 kHz)
Switch off the system
Remove the top cover and, in the case of MOBILETT XP Digital, place it on its side on a
table or similar surface (do not disconnect any cable connections)
Remove the metal plate over board D916 For MOBILETT XP Digital, remove the CXDI
PC (laptop) and also remove the holder plate
Oscilloscope channel 1 at D916 TP “FC2” (26) and D916 TP “GND” (1)
Oscilloscope setting: 1V/div, sweep 5 µs/div
Move switch SW2B on D916 to 2 (service program) and switch the system on
Select service program 7
Use exposure switch S27 to press and hold "PREP"
Set the resonance frequency of the main inverter via potentiometer P4 on D916 to 50
kHz +/-0.5 kHz (Fig. 24 / p. 60)