Siremobil Compact L SPR2-130.841.03 Page 12 of 20 Siemens AG
Rev. 02 06.05 CS PS 24 Medical Solutions
12 Replacement of the 15" LCD Color TFT Display
Geometry 1
Requirements
• The basic unit and the monitor cart are connected to each other;
the system is fully functional.
• The basic unit and monitor cart are connected to each other. The Memoskop is
synchronized by the VIDEOMED DC camera.
• At the MEMOSKOP, call up the SMPTE test pattern. (Technical setup)
• Check the geometry of the displayed test pattern. It should be displayed as shown
in Fig. 4. If adjustments are necessary, follow the description "Manual programming"
(page 16).
• Disconnect the basic unit from the monitor cart (disconnect X10). The MEMOSKOP is in
stand alone mode (internal synchronization).
• Check the geometry of the displayed test pattern. It should be displayed as shown
in Fig. 4. If adjustments are necessary, follow the description "Manual programming"
(page 16).
Fig. 4