SITRANS F M MAGFLO
®®
®®
®
8
SFIDK.PS.026.E5.02
ENGLISH
2.5 Overview and
intrinsically safe data
Specifications:
Supply: 115-230 V or 24 V
Ambient temperature: −20 to 60°C
Enclosure: IP67/
NEMA 4X
2. Installation
SENSOR INTERFACE
1. DN 6 - DN 300 mm
Certificate: Sira 03ATEX1503X
II 2(1)(2) G D
Remote version: EEx d [ia] ia [ib] ib IIB T6 / Compact version: EEx d [ia] [ib] IIB T6
2.5.1 Transmitter data
MAG 6000 I (Ex d)
Electrode
81,82,83,84
IIB IIC
U
o
30 V 30 V
I
o
6.1 mA 6.1 mA
C
o
66 nF 560 nF
L
o
0.96 H 0.96 H
P
o
45.5 mW 45.5 mW
! Coil terminal 85, 86 are “e” terminals.
2. DN 350 - 2000 mm
Certificate: Sira 05ATEX2072X
II 2(1)(2) G D
Remote version: EEx d e [ia] ia [ib] ib IIC T6 / Compact version: EEx d [ia] [ib] IIC T6
User I/O INTERFACE
DN 6 - DN 300
Certificate: Sira 03ATEX1503X
II 2(1)(2) G D
Remote version: EEx d [ia] ia [ib] ib IIB T6 / Compact version: EEx d [ia] [ib] IIB T6
DN 350 - DN 2000
Certificate: Sira 05ATEX2072X
II 2(1)(2) G D
Remote version: EEx d e [ia] ia [ib] ib IIC T6 / Compact version: EEx d [ia] [ib] IIC T6
Electrode Coil
81,82,83,84 85,86
IIB IIC* IIB IIC*
U
o
30 V U
o
30 V
I
o
6.1 mA I
o
105 mA
C
o
560 nF C
o
60 nF
L
o
1 H L
o
2.07 mH
P
o
45.5 mW P
o
3.3 W
* N/A