7ML19985JB02 SITRANS Probe LU (PROFIBUS PA) – INSTRUCTION MANUAL Page 63
mmmmm
G: Data Map
43 Language r/w UNSIGNED_INTEGER (1)
44 Reserved
45 Reserved
46 Device Wear r
Powered Hours UNSIGNED_INTEGER (4)
Poweron Resets UNSIGNED_INTEGER (2)
47 Reserved
48 Reserved
49 Date of Birth r Value ASCII (16)
50
Total Device Operating
Time (seconds)
r Value UNSIGNED_INTEGER (4)
51 Loader Revision r Value ASCII (16)
53
PROFIBUS Current
Consumption
r/w Value INTEGER (1)
54 LCD Contrast r/w Value INTEGER (1)
70
Remaining Device
Lifetime
r
Remaining Device Lifetime UNSIGNED_INTEGER (4)
Status BIT_ENUMERATED (1)
Maintenance ack BIT_ENUMERATED (1)
71 Device Lifetime r/w
Maintenance Required
Limit
UNSIGNED_INTEGER (4)
Maintenance Demanded
Limit
UNSIGNED_INTEGER (4)
Maintenance Alarm
Activation
BIT_ENUMERATED (1)
72
Tota l Expect ed De vi ce
Life
r/w
Total Expect Device Life UNSIGNED_INTEGER (4)
Lifetime Maintenance Ack INTEGER (1)
73
Time Elapsed Since
Last Calibration
r/w Value UNSIGNED_INTEGER (4)
74 Status r
Status BIT_ENUMERATED (1)
Maintenance Calibration
ack
BIT_ENUMERATED (1)
75 Calibration Interval r/w
Maintenance Required
Limit
UNSIGNED_INTEGER (4)
Maintenance Demanded
Limit
UNSIGNED_INTEGER (4)
Maintenance Alarm
Activation
INTEGER (1)
76
Total Calibration
Interval
r/w
Total Calibration Interval UNSIGNED_INTEGER (4)
Calibration acknowledge INTEGER (1)
77 Extended Diagnostics r Value BIT_ENUMERATED (6)
Relative
Index
Name Access Structure Type (cont’d)