Commissioning 
  6.3 Adjustment and initial commissioning 
SIWAREX WL200 
Operating Instructions, 07/2016, A5E02199611B-03 
37 
4.  Calculate the compensating resistance: 
Use the following formula: 
 
R
co
Determined compensation resistance, to be connected in the signal cable SIG+ 
R
Output resistance of the load cells; can also be measured under load 
L
e
Weight error: Differential value from the lowest weight value 
L
t
Test load; placed on all four corners 
N
L
Number of load cells in the scale 
Figure 6-3  Formula for corner load adjustment 
Example calculation: 
 
LC 1:  R
comp
 = R1 = 1004.52 Ω x 12 kg / 
approx. 80 Ω 
LC 2:  - 
1)
  - 
1)
 
LC 3:  R
comp
 = R3 = 1010.70 Ω x 7 kg / 
approx. 47 Ω 
LC 4:  R
comp
 = R4 = 1028.12 Ω x 4 kg / 
approx. 27 Ω 
1)
  Smallest value: No resistance necessary