CAUTION
Looking into laser beam with optical instruments!
Loss of sight possible.
Do not look directly into the laser beam.
5 Adjust the table height so that the body part to be examined is
vertically as near to the gantry isocenter as possible.
6 Align the radiation exposure plane at the body position where you
want to start a scan by fine-tuning the horizontal table position.
7 Switch off the laser light marker.
From this point on, the patient must lie still.
4
Preparing an examination
32 Quick Guide
Print No. HC-C2-015-G.626.08.01.02